High Voltage 50 Ω Pulse Generator for CMTI Application
The CMTI-8010C is a high voltage 50 Ω pulse generator for static and dynamic common mode transfer interference (CMTI) evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17.
Features
- High voltage 50 Ω pulse generator for static and dynamic common mode transfer interference (CMTI) evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17
- Two digital programmable operation modes:
- Mode A: high voltage mode at minimum 0.3 ns pulse rise time
- Mode B: medium voltage mode at minimum 100 ps pulse rise time
- CMTI slew rate:
- up to 10000 V/ns at 3 kV open load output voltage and 0.3 ns rise time (mode A)
- up to 20000 V/ns at 2 kV open load output voltage and 100 ps rise time (mode B)
- 80 kW output power into 50 Ω load (mode A)
- 8 programmable pulse rise times out of: 50 ns, 20 ns, 10 ns, 7 ns, 5 ns, 3.5 ns, 2 ns, 1 ns, 0.6 ns, 0.3 ns, 200 ps, 100 ps (other values on request)
- Optional manual external pulse width extensions from 5 ns to 500 ns (up to 2 µs on request) using the external pulse width extender TLP-8012A5 in mode A up to 3 kV open load output voltage.
- Optional automatic programmable pulse width extensions from 125 ns to 1600 ns (up to 2 µs on request) in 25 ns step size using the external pulse width extender TLP-3011C in mode B up to 2 kV open load output voltage.
- High speed 50 Ω trigger output for oscilloscopes (synchronous to high voltage pulse output)
- Built-in pulse reflection suppression
- Fast pulse repetition rate, typically 100 ms per pulse
- Application programming interface (API): software macros to control pulse generator in standalone operation
- Integrated interlock safety shut-down needed for pulse widths beyond 500 ns
- Industrial isolated and EMI/ESD protected USB control interface
- Two years warranty on the total system
Download: Product Datasheet CMTI-8010C
CMTI-8010C Schematic Front View
CMTI-8010C Schematic Rear View
Accessories suitcase included with CMTI-8010A (typical set)
Item | Sub-Item | Qty. | Descriptions | Model Number |
1 | 1.0 | 1 | High voltage 50 Ω pulse generator for static and dynamic common mode transfer interference evaluations of digital isolators according IEC-60747-5-5 and IEC-60747-17. An interlock safety switch is implemented inside the CMTI-8010C in order to guarantee safe usage of the high voltage capabilities. The customer has to provide adequate measurement set-up housing to guarantee the safety of the operators. | CMTI-8010C |
1.1 | 2 | SMA(m)/SMA(m) adaptor | SMA(m)/SMA(m) | |
1.2 | 2 | SMA(f)/SMA(f) adaptor | SMA(f)/SMA(f) | |
1.3 | 1 | SMA jack T-shape | DELOCK 88711 | |
1.4 | 2 | 50 Ω high performance SMA cable 2m, Sucoflex 104 | SF104-2m | |
1.5 | 1 | Transient voltage suppressor protection for scope input | TVS-70MF-10 | |
1.6 | 1 | 50 Ω SMA Wideband Pick-Off Tee PT-95A | PT-95A | |
1.7 | 1 | High voltage attenuator 50 Ω, 34 dB | HVA-34A | |
1.8 | 1 | High voltage attenuator 50 Ω, 10 dB | HVA-10A | |
1.9 | 1 | High voltage attenuator 50 Ω, 6 dB | HVA-06B | |
1.10 | 1 | SMA torque wrench | 74Z-0-0-21 | |
1.11 | 1 | USB 2.0 cable A male to B male 3 m shielded | 102-1030-BL-F0300 | |
1.12 | 1 | Power line cable according supply voltage, 1.5 m | PLC-150A | |
1.13 | 1 | Hardware manual CMTI-8010C | CMTI-HM | |
1.14 | 1 | Safety interlock shut-down to allow 2 µs pulse width (optional) | CMTI-IN-LO |